An issue with the plunger rod seal was identified where the seal had the potential to migrate inside the pump case top during slow rate/ high back pressure infusions.
Source facts
Action
On May 14th, 2010, Smiths Medical contacted the facility that received the pumps via e-mail and requested that they immediately return the pumps. The pumps were returned per Returned Goods Authorization (RGA) and exchanged for replacement pumps manufactured with plunger rod seal G6000077.
Initiated
2010-05-14
Root cause
Device Design
Terminated
2013-04-11
Distribution
Distributed in the state of NH.
Recall number
Z-0996-2013
Recall status
Terminated
CDRH record ID
114372
FDA classified
2013-03-26
Recall event number
63621
Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.
Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.