Difficulty inserting an obturator and/or suction catheter into the tracheostomy tube.
Source facts
Action
Consignees were sent an "Urgent Device Alert" letter dated 01/14/2009, by mail via Federal Express overnight delivery and instructed to evaluate any of the affected lot numbers of tracheostomy tubes prior to placement into a patient's stoma. If either the obturator or the desired sterile suction catheter does not fit properly into the tube, the tube must not be inserted into the patient. Further instructions are available at Nellcor Technical Services by calling 1-800-635-5267, option 3, then option 1. New product was expected by mid February, 2009, to address the issue.
Initiated
2009-01-14
Root cause
Finished device change control
Terminated
2012-05-21
Distribution
Nationwide Distribution, including Puerto Rico.
Recall number
Z-0997-2009
Recall status
Terminated
CDRH record ID
78880
FDA classified
2009-03-04
Recall event number
51160