Z-0997-2026: BECKMAN COULTER BICARBONATE REF: OSR6137 OSR6237 OSR6637 OSR6x37 Bicarbonate is a system reagent for the

Recall record Z-0997-2026 for Beckman Coulter Ireland, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0997-2026
Company in FDA record
Beckman Coulter Ireland, Inc.
FDA classification date
Dec 23, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Bicarbonate reagent may generate falsely high Bicarbonate results due to interference with Lactate Dehydrogenase (LDH) in the test sample.

Source facts

Action
On 10/14/2025, the firm sent via US Postal and email an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that complaints have been received for elevated results generated using Bicarbonate reagent when there are high levels of lactate Dehydrogenase (LDH) in the test samples. Beckman Coulter has investigated and confirmed that the lots listed above of Bicarbonate reagent may generate Bicarbonate results that are biased high due to interference with LDH in the test sample. Customers are instructed to: "Discontinue use and dispose all Bicarbonate OSR6x37 lots listed in Table A in the ISSUE section of this letter, following their laboratory protocols and local regulations. Contact their local Beckman Coulter representative for replacement or re-imbursement. "For Bicarbonate OSR6x37 lots listed in Table B in the ISSUE section if elevated LDH levels are suspected, please assess results in conjunction with patient s LDH values. "Share the content of this letter with their laboratory and/or Medical Director to evaluate the requirement for a retrospective review of Bicarbonate results for neonatal patients and patients with elevated LDH levels. o Complete the attached Replacement Order Form and email to askbeckman@beckman.com or fax to (866) 294-7850 OR o Call Client Services at (800) 526-3821 For questions, contact Customer Support Center. "From Beckman Coulter's website: http://www.beckmancoulter.com "By phone: call 1-800-854-3633 in the United States.
Initiated
2025-10-14
Root cause
Nonconforming Material/Component
Distribution
Worldwide - U.S. Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, Guan, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. The countries of Argentina, Brazil, Canada, Ecuador, Ireland, Mexico, Netherlands, Panama, Taiwan, and Province of China.
Recall number
Z-0997-2026
Recall status
Open, Classified
CDRH record ID
216706
FDA classified
2025-12-23
Recall event number
97842

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗