Skip to content

Public FDA record

Z-0999-2013: CareStream Dental CS 2100 X-ray system Product Usage: The expected usage (intended use) for all of these

Recall record Z-0999-2013 for Trophy Sas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0999-2013
Company in FDA record
Trophy Sas
FDA classification date
May 2, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There have been approximately 2% of the x-ray generators manufactured between February 2012 and November 2012 returned due to errors. These generators showed the type of defect that indicates poor contact of HT transformer power pin. The defect is a bent connector associated with the generator.

Source facts

Action
Trophy, A Carestream Dental Company, - A Subsidiary of Carestream Health Inc. The Center for Devices (CDRH) understands that the corrective action plan (CAP) involving 167 units will consist of implementing rework procedure TL008, which describes the procedure for a bending the metallic strip the female connector to insure strong connection with the male counterpart. The CAP appears to adequately address the problem and is hereby approved For further questions please call (585) 781-1997.
Initiated
2012-09-18
Root cause
Radiation Control for Health and Safety Act
Terminated
2017-05-02
Distribution
USA Nationwide Distribution
Recall number
Z-0999-2013
Recall status
Terminated
CDRH record ID
116555
FDA classified
2013-05-02
Recall event number
64589

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗