Skip to content

Public FDA record

Z-0999-2022: da Vinici X da Vinci Xi Sureform 45 REF 480445 da Vinci X da Vinci Xi Sureform 45 Curved-Tip REF 48054

Recall record Z-0999-2022 for Intuitive Surgical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0999-2022
Company in FDA record
Intuitive Surgical, Inc.
FDA classification date
Apr 22, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

As a result of complaint data review, it identified tissue pushout events occurring during firing of the SureForm stapler when tissue is pushed distally within the jaws, rather than being held in place and transected. Overall rate of harm associated with bleeding of critical severity exceeds the predicted likelihood of harm as documented within Intuitive s existing risk documentation.

Source facts

Action
On 03/17/2022, the firm sent an Urgent Medical Device Correction letter informing customers of a potential for injury associated with the stapler instruments if target tissue fails to remain in place with the jaws of the stapler during the stapler firing sequence. This phenomenon is referred to as a tissue pushout event. As the staple firing sequence progresses, the target tissue is transected but the target tissue is not approximated due to lack of staple engagement with the tissue. This type of event may result in un-approximated tissue and occurs when attempting to cross over existing staple lines in the creation of a continuous staple line. Place this letter with the User Manual. Notify all surgeons and personnel using the Surgical System to review and understand contents of this letter and reacquaint themselves by reading the instructions, warnings, and cautions in the SureForm Instruments and Accessories User Manual Addendum. Customers can continue to use their instruments by following instructions: Use caution when unable to avoid crossing over existing staple lines and continue to adhere to all existing warnings and cautions found in the SureForm Instruments and Accessories User Manual Addendum. If a tissue pushout event is occurring, users may limit unapproximated tissue by pressing the emergency stop button (on Surgeon Console or Patient Cart helm), after which the SureForm stapler can be safely unclamped from the tissue using the Manual Release Knob. For further assistance, contact Clinical Sales Representative or Customer Service at: -US, Puerto Rico and Dominican Republic: (800) 876-1310, Option 3 (4AM to 5PM PST) or mail: customerservice@intusurg.com -Europe, Middle East, Asia, South America and Africa: +800 0821 2020 or +4121 821 2020 (8AM to 6PM CET) or EUCS@intusurg.com -South Korea: 02-3271-3200 (9AM to 6PM KSTJ) or support.korea@intusurg.com
Initiated
2022-03-17
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - U.S. Nationwide distribution in the states of AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, Puerto Rico, RI, SC, SD, TN, TX, UT, VT, VA, WA, WV, WI, and WY. The countries of Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Lebanon, Luxembourg, Malaysia, Malta, Martinique, Monaco, Netherlands, New Zealand, Norway, Panama, Poland, Portugal, Qatar, Reunion, Romania, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Recall number
Z-0999-2022
Recall status
Open, Classified
CDRH record ID
192324
FDA classified
2022-04-22
Recall event number
89736

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗