Skip to content

Public FDA record

Z-0999-2026: LUMINOS Lotus Max, Model 11574100 and Luminos dRF Max, Model 10762471

Recall record Z-0999-2026 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-0999-2026
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jan 8, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A potential issue associated with LUMINOS Lotus Max and Luminos dRF Max systems with software version VF11 with bucky wall stand for mobile detectors and without ceiling stand .When switching quickly between organ programs (OGPs) the collimation may be incorrect.

Source facts

Action
The Customer Advisory Notice (CAN) was sent to the affected customer on December 5, 2025. The CAN notifies the customer of the problem and provides steps the user should take to avoid the problem until the resolution software update is available. When switching within five seconds from a table related OGP to a bucky wall stand related OGP with top-alignment and then back to a table related OGP, it may happen, that the collimation is not adapted to the limitations of the detector. The active radiation field may thus extend beyond the limits of the detector, resulting in radiation outside of the area of the detector.
Initiated
2025-12-05
Root cause
Radiation Control for Health and Safety Act
Distribution
U.S. and OUS
Recall number
Z-0999-2026
Recall status
Open, Classified
CDRH record ID
217937
FDA classified
2026-01-08
Recall event number
98230

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗