Reason stated in the source
Defibrillator may fail to charge or hold its charge above energies of 75 joules
Source facts
- Action
- Zoll Medical issued a letter on 6/25/03 to domestic consignees and 6/26/03 to international consignees on 6/26/03. Letters were issued Certified Mail Return Receipt requested. Instructions are provided for Biomedical to charge device and then dump the charge. A damaged device will fail to charge and issue an error message. Users then are requested to return device. Firm will initiate telephone contact on 6/30/03 to all accounts to verify receipt of notice and confirmation of testing per the instructions.
- Initiated
- 2003-06-23
- Root cause
- Other
- Terminated
- 2020-12-29
- Distribution
- Nationwide Foreign: Emirates, Australia, Brussels, Brazil, Canada, China, Columbia, Chezche, Germany, Spain, Finland, France, Greece, Hong Kong, India, Indonesia, Italy, Japan, Korea, Kuwait, Latvia, Mexico, Netherlands, Poland,Romania, Saudi Arabia, Syrai, Thailand, Turkey,