Skip to content

Public FDA record

Z-1000-2022: NeuMoDx Cartridge-For in vitro diagnostic use with NeuMoOx Molecular Systems with either of the followin

Recall record Z-1000-2022 for NeuMoDx Molecular Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1000-2022
Company in FDA record
NeuMoDx Molecular Inc
FDA record creation date
Apr 22, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential for false positive results for SARS-CoV-2 in identified lots of cartridges, when used in conjunction with assays containing SARS-CoV-2 targets (NeuMoDx" SARS-CoV-2 Test Strip Cat #300800 and NeuMoDx" Flu A-B/RSV/SARS-CoV-2 Vantage Assay Cat #300900) result incorrect management of patients

Source facts

Action
QIAGEN issued customer Urgent Medical Device Correction notification via email firstly by English (then in local language if required)on 2/18/22. Letter states reason for recall, health risk and action to take: 1. IMMEDIATELY discontinue use of the product and discard remaining inventory in accordance with your national and local safety and environmental regulations. Please contact QIAGEN Technical Service for a free-of-charge replacement of scrapped inventory. 2. Share this Field Safety Corrective Action with all users of the NeuMoDx Cartridges in your facility to ensure they are aware of this notice. 3. If you have already used NeuMoDx Cartridges from any of these lots in combination with the NeuMoDx SARS-CoV-2 Assay, the NeuMoDx Flu-A-B/RSV/SARS-CoV-2 Assay and NeuMoDx Laboratory Developed Tests (LDTs) for SARS-CoV-2 RNA detection please review all SARS-CoV-2 positive results to exclude erroneous diagnosis and treatment, except in those cases where alternative confirmation was obtained. Note: results for targets other than SARS-CoV-2 are not impacted by this Urgent Field Safety Corrective Action. 4. Complete the attached response form and return to Quality.Communications@qiagen.com to confirm receipt of this notification. If you have any questions regarding this notice, please contact QIAGEN Technical Services at 1-800-362-7737 or techservice-na@qiagen.com
Initiated
2022-02-18
Root cause
Material/Component Contamination
Distribution
Worldwide distribution - US Nationwide distribution in the states of OH, MO, VA, MI, MI, FL, MD, IA, NY, MO and the countries of Italy, Germany, Switzerland, Sweden, Netherlands, Finland, Spain, United Kingdom, Belgium, Austria, Hong Kong, France, Denmark, Slovenia.
Recall number
Z-1000-2022
Recall status
Open, Classified
CDRH record ID
192674
FDA record created
2022-04-22
Recall event number
89846

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗