Skip to content

Public FDA record

Z-1000-2024: HRIS ACET CUP CUT TIP 32X140 Part Number: 6210-5-200rker. single-use, sterile surgical instruments use

Recall record Z-1000-2024 for Howmedica Osteonics Corp.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1000-2024
Company in FDA record
Howmedica Osteonics Corp.
FDA classification date
Feb 6, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Failed to meet the acceptance criteria for the seal integrity and/or package integrity testing for the Sterile Barrier Outer Pouch and the Protective Barrier Inner Pouch, compromising product sterility

Source facts

Action
Stryker issued Urgent Medical Device Recall (UMDR) PFA 3285098 was issued by third party Sedgwick on 22/DEC/2023 via UPS 2-day air. OUS customer notification will be issued at the country level. Letter states reason for recall, health risk and action to take: 1. Please inform users of this Urgent Medical Device Correction and forward this notice to all individuals who need to be made aware. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Dispose of all affected products listed in the Part and Lot Number Attachment (PFA 3285098). 4. Complete and sign the enclosed Urgent Medical Device Correction Business Reply Form. " Medical Facility: Please sign the Business Reply Form and email to: strykerortho4742@sedgwick.com / Fax: (866) 723-2482 " Stryker Branch/Agency: please sign the AdobeSign form arriving to you via email. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. Please assist us in meeting our regulatory obligation by emailing back the attached Urgent Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physical inventory on site. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (888)-756-7846. For questions pertaining to the correction, email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com
Initiated
2023-12-22
Root cause
Package design/selection
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Brazil, Canada, Chile, China (Hong Kong), Columbia, Italy, Japan, Malaysia, Netherlands (Venlo), Slovenia, S. Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom.
Recall number
Z-1000-2024
Recall status
Open, Classified
CDRH record ID
205598
FDA classified
2024-02-06
Recall event number
93830

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗