A mix-up of autosampler tube positions with a possibility of misidentification of sample results and a risk of operator injury from an exposed aspiration needle.
Source facts
Action
Consignees notified via email and postal delivery with recall letter dated 01/02/2020. The letter identified affected product, stated reason for recall, and provided specific actions to follow. A Recall Acknowledgement Form was requested to be returned by email, postal delivery, or fax.
Clinical Diagnostic Solutions Inc. is handling US consignees only.
Boule Medical AB is handling OUS consignees.
Initiated
2020-01-02
Root cause
Software design
Terminated
2022-11-03
Distribution
Worldwide distribution. US Nationwide, Venezuela, Africa, Italy, Brazil, and Ukraine. No governmental, Canadian, or Mexican consignees.
Recall number
Z-1001-2020
Recall status
Terminated
CDRH record ID
178758
FDA record created
2020-02-05
Recall event number
84614