Z-1003-2013: Philips Healthcare Computed Tomography X-Ray System. These devices are whole-body computed tomography (C

Recall record Z-1003-2013 for Philips Healthcare Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1003-2013
Company in FDA record
Philips Healthcare Inc.
FDA classification date
Apr 17, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.

Source facts

Action
Philips Healthcare proposed corrective action plan (CAP) is comprised of the following elements: * Updating procedures in manufacturing to verify, either through software calibration or through gage-pin measurement, that A-plane collimation is appropriate to the design specification. It is our understanding that these procedures have been implemented. * For refurbished PET/CT and SPECT/CT systems distributed by Philips, the manufacturing process now requires that CTDI testing be conducted in addition to air testing and that each system meet updated CTDI tolerance limits to be published in the product's instructions for use. * Reassessing and correcting the correlation between air dose and CTDI specifications for all CT systems or sub-systems, resulting in updates of the CTDI specifications (tolerances) to be published in the instructions for use. It is our understanding that this reassessment and correction have been implemented, that the updated CTDI specifications will be made available as new CTDI tolerance tables with the customer information letter to be distributed immediately, and that these updated CTDI specifications will also be distributed as an addendum to the instructions for use manuals to the installed base during the next 12 months. * For all Brilliance CT, Gemini PET/CT, and Precedence SPECT/CT systems, in the installed base as well as in forward production, instructions for field service will be updated to provide clear directions on how to verify that the A-plane collimator has been appropriately adjusted to ensure expected product performance after its replacement and/or the replacement of an x-ray tube. This plan appears to adequately address the problem, and it is hereby approved. For further questions please call (978) 687-1501.
Initiated
2010-04-05
Root cause
Radiation Control for Health and Safety Act
Terminated
2021-03-08
Distribution
Worldwide Distribution including USA (nationwide)
Recall number
Z-1003-2013
Recall status
Terminated
CDRH record ID
116837
FDA classified
2013-04-17
Recall event number
64708

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗