Open original FDA record ↗Recall record
- FDA identifier
- Z-1003-2022
- Company in FDA record
- Boston Scientific Corporation
- FDA record creation date
- Apr 25, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Users need to be aware that embolism is a possible adverse event associated with hydrogel misplacements. Updated IFU provides new procedural instructions, warnings and precautions that describe steps to verify correct placement of the hydrogel following implantation, as well as technique recommendations for the proper placement of SpaceOAR and SpaceOAR Vue.
Source facts
- Action
- An URGENT MEDICAL DEVICE PRODUCT ADVISORY dated 2/24/22 was sent to customers. Inadvertent placement of SpaceOAR gel into a blood vessel and subsequent migration of the hydrogel outside of the pelvis, led to embolism. Boston Scientific will incorporate best practices regarding hydrogel placement during all SpaceOAR procedures into the IFU. The updated IFUs will be packaged with SpaceOAR and SpaceOAR Vue Systems after regulatory approvals are obtained for the IFU updates. This Product Advisory contains the recommended IFU updates, located in Appendix 1, intended to: -Raise awareness of embolism as an existing adverse event. -Add new Warnings and Precautions that describe technique recommendations for the placement of SpaceOAR and SpaceOAR Vue to reduce the unintentional placement of hydrogel. - Add procedural instructions to reduce the unintentional placement of hydrogel in a blood vessel and steps to verify placement of the hydrogel following implantation. NO product is being recalled and you are NOT required to return product to Boston Scientific. As per the IFU, SpaceOAR and SpaceOAR Vue System procedures should only be performed by physicians who have received appropriate training for proper hydrogel spacer deployment technique. The Acknowledgement Form must be completed and returned even if you do not have any affected units. Return the Acknowledgement Form by Email: BSCFieldActionCenter@bsci.com or Fax to: Field Action Center 1-763-415-7708
- Initiated
- 2022-02-24
- Root cause
- Labeling design
- Distribution
- Worldwide - US Nationwide distribution.
- Recall number
- Z-1003-2022
- Recall status
- Open, Classified
- CDRH record ID
- 192708
- FDA record created
- 2022-04-25
- Recall event number
- 89862