Skip to content

Public FDA record

Z-1003-2023: Software : myNeedle Guide 2D license, myNeedle Guide 2D AWP only Workflow license, or myNeedle Guide 3D l

Recall record Z-1003-2023 for Siemens Medical Solutions USA, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1003-2023
Company in FDA record
Siemens Medical Solutions USA, Inc
FDA classification date
Jan 20, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

In the syngo.CT VA40 and syngo.CT VA50 myNeedle Guide application, the order of the orientation labels head , center, and feet in the axial thick slice segments are erroneously swapped

Source facts

Action
Siemens Customer Safety Advisory Notification CT083/22/S issued beginning on 12/23/2022 via email. Letter states reason for recall, health risk and action to take: When performing a CT intervention study, please make sure that the Recon parameter Image order is set to Head to Feet. Provides instruction on how to avoid the issue. Siemens is currently working on a permanent solution for this issue. This solution will be rolled out as a software update to all affected customers free of charge. Siemens is currently working on a permanent solution for this malfunction. This solution will be rolled out as a software update to your CT Scanner. The Siemens Service Organization will contact you as soon as the update is available. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understood the content therein. Regional Contact: Name: Rebecca Tudor Department: Post Market Regulatory Compliance Telephone: (484) 323-4198 Email: rebecca.tudor@siemens-healthineers.com
Initiated
2022-12-22
Root cause
Under Investigation by firm
Distribution
Nationwide
Recall number
Z-1003-2023
Recall status
Open, Classified
CDRH record ID
197992
FDA classified
2023-01-20
Recall event number
91481

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗