Z-1003-2025: Brand Name: Olympus Thunderbeat Front-Actuated Grip Type S Hand Instrument Product Name: Thunderbeat, 5MM

Recall record Z-1003-2025 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1003-2025
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Jan 21, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The probe tips of the hand instruments are being damaged or breaking including pad damage and detachment.

Source facts

Action
On December 20, 2024, URGENT FIELD SAFETY NOTICE: ADVISORY letters were sent to customers. Actions Required: Olympus requires you to take the following actions: 1. Carefully read the content of this FSN. 2. Ensure all personnel are completely knowledgeable and thoroughly trained on the content of this FSN and the relevant IFU. 3. Olympus requests that you acknowledge receipt of this FSN through our web portal: a. Go to https://olympusamerica.com/recall b. Enter the recall number "0462 c. Complete the form as instructed and include your account ID number: [XXXXXXXX] d. Indicating in the comments if you would like to receive in-service training to support the correct use of this device. 4. If you have further distributed this product, identify your customers, and forward this FSN to them. Olympus requests that you report any complaints, including probe damage or pad detachment, to Technical Assistance Center (TAC) at 1- 800-848-9024 (option 1), and the FDA. Adverse events experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Event Reporting program either online, by regular mail, or by fax. Olympus fully appreciates your prompt cooperation. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 or by e-mail at Cynthia.Ow@olympus.com.
Initiated
2024-11-25
Root cause
Under Investigation by firm
Distribution
US Nationwide distribution and International: Algeria, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Belize, Benin, Bolivia, Bosnia-Herz., Brazil, Bulgaria, Burkina Faso, Cambodia, Canada, Chile, China, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Estonia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Guernsey, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, Kuwait, Latvia, Lithuania, Luxembourg, Malaysia, Mauritius, Mexico, Montenegro, Morocco, Nepale, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Thailand, Tunisia, Turkey, Ukraine, United Kingdom, Uruguay, Utd.Arab Emir., Vietnam, Yemen.
Recall number
Z-1003-2025
Recall status
Open, Classified
CDRH record ID
212142
FDA classified
2025-01-21
Recall event number
95950

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗