Open original FDA record ↗Recall record
- FDA identifier
- Z-1004-2011
- Company in FDA record
- Optovue Inc.
- FDA classification date
- Feb 1, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Use of the CAM attachment may result in a positioning anomaly and mistreatment in the eyes of the patient.
Source facts
- Action
- Optovue Inc. sent a Field Modification/Correction Notice dated July 20, 2010, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were informed that they could continue to use the iVue until the correction has been made. Customers were instructed to review any existing iVue patient data to ensure that this situation and an improper designation of eye assignment has not occurred. Upon completion of the updates, customers were to sign the attached form indicating they acknowledge that their iVue has been updated and they understand that they must review existing data. For any questions regarding this recall call 1-866-344-8948 or by mail at 45531 Northport Loop W, Fremont, CA 94538.
- Initiated
- 2010-07-20
- Root cause
- Software design (manufacturing process)
- Terminated
- 2011-02-08
- Distribution
- Nationwide
- Recall number
- Z-1004-2011
- Recall status
- Terminated
- CDRH record ID
- 94463
- FDA classified
- 2011-02-01
- Recall event number
- 56759