Open original FDA record ↗Recall record
- FDA identifier
- Z-1004-2015
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Jan 20, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2.
Source facts
- Action
- Ortho Clinical Diagnostics sent an Important Product Correction Notification letter dated September 29, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed Calibration may not occur when using calibrator barcode labels supplied with VITROS Chemisty Products Calibrator Kit 2. The resolution to this software anomaly will be in a future version of software. Until the new software is installed on their VITROS System, customers were instructed not to use the calibrator barcode labels for VITROS Calibrator Kit 2. Customers were advised that you manually program the calibration. Customers with questions were instructed to contact Customer Technical Services at 1-800-421-3311.
- Initiated
- 2014-09-29
- Root cause
- Software Manufacturing/Software Deployment
- Terminated
- 2017-04-03
- Distribution
- Worldwide Distribution - USA including CO, NC, NJ, NY, PA, MI, WV and Internationally to Canada, Japan and Spain.
- Recall number
- Z-1004-2015
- Recall status
- Terminated
- CDRH record ID
- 131433
- FDA classified
- 2015-01-20
- Recall event number
- 69727