Potential for device to shut down prematurely under certain circumstances.
Source facts
Action
On 4/13/07 the firm released Service Bulletin part no. 810-2614-00, SB - AED 10 S/W V020600 DOWNGRADE to V020500. That mandatory bulletin, for units with software version 2.06.00 and up indicates that all returning AED 10 units shall have their software version downgraded to 2.05.00.
Initiated
2008-12-15
Root cause
Device Design
Terminated
2012-02-15
Distribution
Worldwide distribution: USA, Belgium, Canada, France, Ireland, Israel, Mexico, and United Kingdom.
Recall number
Z-1005-2009
Recall status
Terminated
CDRH record ID
74811
FDA classified
2009-03-10
Recall event number
50142