Z-1006-2018: ExacTrac versions 6.5 through 6.5 intended to be used to place patients at an accurately defined point wi

Recall record Z-1006-2018 for Brainlab AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1006-2018
Company in FDA record
Brainlab AG
FDA record creation date
Mar 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The usage of workflows that deviate from the recommended specifications in the User manual for ExacTrac Patient Positioning System (versions 6.0, 6.1, 6.2, 6.5) with Auxiliary Device Interface (ADI) and Varian Clinac or Varian TrueBeam, which may result in misinterpretation of beam authorization via ADI.

Source facts

Action
The firm disseminated their notices by email on 02/14/2018. The notice requests the following actions: "User Corrective Action: Always follow the instructions below during use of ExacTrac Patient Positioning System (versions 6.0, 6.1, 6.2, 6.5) with Auxiliary Device Interface (ADI) and Varian Clinac or Varian TrueBeam: Do not use beam authorization by ExacTrac via ADI as a safety feature to ensure correct patient positioning before treatment. For any treatments performed with ExacTrac always verify that ExacTrac confirms the target position by displaying the OK icon. Do not treat the patient if the OK icon is not displayed. This indicates that the isocenter is in the planned treatment position. The treatment beam should not be activated unless the OK icon is shown. If using ExacTrac on a Varian Clinac or Varian TrueBeam system equipped with the Auxiliary Device Interface (ADI) always perform treatment in the following sequence: 1. Start the ExacTrac Software and login using your user name and password 2. Load the Plan / Patient on the Varian Treatment Application / Console 3. Open the Plan / Patient on ExacTrac by clicking Open in the message box (refer to Figure 2) 4. Prepare the beam on the Varian Treatment Application / Console If you intend to perform treatments for which ExacTrac is not used for patient positioning, please click Cancel in the message box to confirm not opening the patient and to continue the treatment without ExacTrac. ExacTrac can be configured to directly load the patient automatically in ExacTrac if it is opened on the Varian Treatment Application / Console - meaning that the message box (refer to Figure 2) is deactivated. However, this may not be the preferred solution for each user, as automatic loading causes delays in treatment preparation if no positioning via ExacTrac is to be used, since in these cases the patient data has to be closed manually. Please contact your local Brainlab Customer Support Representat
Initiated
2018-02-14
Root cause
Other
Terminated
2021-05-17
Distribution
worldwide
Recall number
Z-1006-2018
Recall status
Terminated
CDRH record ID
161795
FDA record created
2018-03-14
Recall event number
79266

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗