Open original FDA record ↗Recall record
- FDA identifier
- Z-1006-2024
- Company in FDA record
- Ivoclar Vivadent, Inc.
- FDA classification date
- Feb 7, 2024
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Affected product may not perform as expected; the amalgam powder may not triturate correctly and the physical properties (i.e. strength, compression, working time, etc.) will be adversely affected.
Source facts
- Action
- On December 19, 2023, the firm notified customers by phone. On December 21, 2023, the firm further notified affected customers via URGENT MEDICAL DEVICE RECALL letters through email. Customers were instructed to check their inventory for the affected lot and quarantine the product, and stop deliveries of the affected lot. Customers should respond to the letter and return all stock of affected product to the recalling firm. If product was further distributed, the customer should provide downstream customers' information to the recalling firm so that the firm may communicate recall information to them directly. Firm stated that product manufactured after December 6, 2023, were corrected for any irregularities. Customers with questions about risk minimizing safety measures may contact customercare.vigilance@ivoclar.com. For questions regarding replacement or credit note for products affected by this field safety notice, please contact customer service at Ivoclar Vivadent Inc at 1-800-533-6825.
- Initiated
- 2023-12-21
- Root cause
- Employee error
- Distribution
- US Nationwide distribution in the states IN and TN.
- Recall number
- Z-1006-2024
- Recall status
- Open, Classified
- CDRH record ID
- 205447
- FDA classified
- 2024-02-07
- Recall event number
- 93796