Open original FDA record ↗Recall record
- FDA identifier
- Z-1007-2017
- Company in FDA record
- Medtronic Navigation, Inc.
- FDA classification date
- Jan 11, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Medtronic Navigation, Inc. announces a voluntary field action for the AxIEM Touch N Go Pointer Probe because the patient registration accuracy may compromise navigation accuracy when using the StealthStation System with Touch N Go Pointer Probes for patient registration.
Source facts
- Action
- The firm, Medtronic, sent an "URGENT: MEDICAL DEVICE RECALL" letter on 12/12/16 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to examine inventory for any affected products, quarantine them for return to Medtronic; complete and return the CUSTOMER RESPONSE FORM by email at RS.NavFCA@medtronic.com or fax to Medtronic Technical Services at 651-367-7075 (even if you do not have affected product); and contact Medtronic Technical Services at 1-800-595-9709 to get a return material authorization (RMA) and arrange for their return and no charge replacements. If you have any questions regarding this communication, please contact Medtronic Technical Services at (800) 595-9709.
- Initiated
- 2016-12-12
- Root cause
- Process change control
- Terminated
- 2017-05-19
- Distribution
- US Nationwide Distribution.
- Recall number
- Z-1007-2017
- Recall status
- Terminated
- CDRH record ID
- 152061
- FDA classified
- 2017-01-11
- Recall event number
- 76060