Medtronic is voluntarily recalling one lot of its
MAHURKAR"* Elite 12 FR x 16cm Acute Dual Lumen Dialysis Catheter with Curved Extensions, IC Trays, Lot 1717900160, following notification from our supplier that one production lot of MAHURKAR"* Elite 12 FR x 16cm Acute Dual Lumen Dialysis Catheter with Curved Extensions, IC Trays was inadvertently shipped to Medtronic and distributed without being sterilized.
Source facts
Action
Medtronic began notifying customers via telephone on October
14, 2017, and all consignees were notified via Federal Express or
certified mail on October 17, 2017. The letter informs customers
that they were shipped devices that were not sterile. Customers
are requested to acknowledge receipt and understanding of the
Urgent Field Corrective Action Notice.
Initiated
2017-10-14
Root cause
Process control
Terminated
2018-06-15
Distribution
U.S.
Recall number
Z-1007-2018
Recall status
Terminated
CDRH record ID
161871
FDA record created
2018-03-14
Recall event number
79319