Open original FDA record ↗Recall record
- FDA identifier
- Z-1007-2022
- Company in FDA record
- Carl Zeiss Meditec AG
- FDA record creation date
- Apr 26, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
High friction of the slider can cause the device to stick, or not move as intended.
Source facts
- Action
- On 03/16/2022, the Recalling firm called U.S. Customers to inform them of that is has been determined that single lot (Lot FG21082410) of the ZEISS miLoop may have been provided with potentially too high friction of the slider. A follow-up email was sent on 03/22/2022 that contained an "Urgent Medical Device Recall" letter. Customer are asked to discontinue the use and not to further distribute the specific product lot, and to fill out the attached acknowledgement form to dl.med-qualitycti.us@zeiss.com. Customer are asked to return all devices from lot FG21082410 to Carl Zeiss Meditec AG. For question or assistance, contact: Scott Smith, Director of Sales at (704) 906-3287 or email scott.smith@zeiss.com
- Initiated
- 2022-03-16
- Root cause
- Under Investigation by firm
- Distribution
- U.S.: AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OR, PA, Puerto Rico, SC, TN, TX. UT, VA, WA, and WV O.U.S.: Not provided
- Recall number
- Z-1007-2022
- Recall status
- Open, Classified
- CDRH record ID
- 192932
- FDA record created
- 2022-04-26
- Recall event number
- 89928