Skip to content

Public FDA record

Z-1007-2022: ZEISS miLOOP Lens Fragmentation Device REF 303071-9090-000. Used during cataract surgery.

Recall record Z-1007-2022 for Carl Zeiss Meditec AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1007-2022
Company in FDA record
Carl Zeiss Meditec AG
FDA record creation date
Apr 26, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

High friction of the slider can cause the device to stick, or not move as intended.

Source facts

Action
On 03/16/2022, the Recalling firm called U.S. Customers to inform them of that is has been determined that single lot (Lot FG21082410) of the ZEISS miLoop may have been provided with potentially too high friction of the slider. A follow-up email was sent on 03/22/2022 that contained an "Urgent Medical Device Recall" letter. Customer are asked to discontinue the use and not to further distribute the specific product lot, and to fill out the attached acknowledgement form to dl.med-qualitycti.us@zeiss.com. Customer are asked to return all devices from lot FG21082410 to Carl Zeiss Meditec AG. For question or assistance, contact: Scott Smith, Director of Sales at (704) 906-3287 or email scott.smith@zeiss.com
Initiated
2022-03-16
Root cause
Under Investigation by firm
Distribution
U.S.: AL, AR, AZ, CA, CT, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, MT, NC, NJ, NY, OH, OR, PA, Puerto Rico, SC, TN, TX. UT, VA, WA, and WV O.U.S.: Not provided
Recall number
Z-1007-2022
Recall status
Open, Classified
CDRH record ID
192932
FDA record created
2022-04-26
Recall event number
89928

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.