Weld protrusion/physical gap between the housing and phaco pack cover, may exceed design specification, which could lead to priming cycle failure and/or suboptimal vacuum delivered to the phacoemulsification and irrigation/aspiration handpieces during surgery, which could result in surgical delay and/or longer surgical time, which could result in transient corneal edema.
Source facts
Action
On 12/07/22, recall notices were sent to customers who were asked to do the following:
1) Discontinue using affected product.
2) Return product to the recalling firm.
3) Complete and return the Customer Reply form via email to RegCompliOne@its.jnj.com
4) Distributors, distribute the recall notice to your customers.
Customers with questions or concerns can call the firm at 1-877-266-4543 Option 2.
Initiated
2022-12-07
Root cause
Process control
Distribution
US: FL, MS, NY, TN, TX, IL, MN, OH.
OUS: Belgium, China, Estonia, France, Germany, Indonesia, Italy, Japan, Malaysia, Netherlands, Portugal, Sri Lanka, Switzerland, Thailand, United Kingdom
Recall number
Z-1007-2023
Recall status
Open, Classified
CDRH record ID
197506
FDA classified
2023-01-20
Recall event number
91290