Z-1007-2024: ExcelsiusGPS Cannula A, 9mm x 250mm, Cannula used with ball tip probes, Model No. 6143.2844

Recall record Z-1007-2024 for Globus Medical, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1007-2024
Company in FDA record
Globus Medical, Inc.
FDA classification date
Feb 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Devices may not be rendered sterile due to an internal gap that may be present on some devices causing increased difficulty cleaning prior to steam sterilization. There is an increased risk for infection, inflammation, local toxicity, systemic toxicity, sensitivity, and cross-contamination.

Source facts

Action
On December 15, 2023, the firm began to notify customers by phone. On January 9, 2024, firm began following up with letters. Customers were instructed to immediately return affected cannulas to Globus Medical and complete the letter's Response Card. For recall-related questions, please contact Globus Medical Complaint Handling Unit at recall@globusmedical.com. Fax: 610-300-1342.
Initiated
2023-12-15
Root cause
Device Design
Distribution
Worldwide - US Nationwide and the countries of Australia, Austria, Belgium, Czech Republic, Germany, India, Indonesia, Ireland, Italy, Japan, Malaysia, Netherlands, Poland, Saudi Arabia, Singapore, Spain, United Arab Emirates, United Kingdom.
Recall number
Z-1007-2024
Recall status
Open, Classified
CDRH record ID
205465
FDA classified
2024-02-07
Recall event number
93804

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗