Open original FDA record ↗Recall record
- FDA identifier
- Z-1008-2011
- Company in FDA record
- Accuray Inc
- FDA classification date
- Feb 1, 2011
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
If electron density values are left empty, the calculation of radiation dose in a patient will be modeled as air-like density material rather than correct density. A plan may be created and saved, thus creating the risk of mistreatment.
Source facts
- Action
- The firm, Accuray, sent an "URGENT DEVICE CORRECTION" letter dated December 29, 2009, to all customers. The letter described the product, problem, and action to be taken by the customers. The customers were instructed to verify correctness of the CT number to relative electron density and Mass Density Calibration curves. Note: A patch is being developed to address this issue. Accuray Customer Support will contact the customers regarding further resolution. If you have any questions or concerns regarding this issue, please contact Accuray Customer Support at +1-877-668-8667 (USA) or +1-408-716-4700 (non USA) or by email: customersupport@accuray.com.
- Initiated
- 2009-12-29
- Root cause
- Employee error
- Terminated
- 2011-02-11
- Distribution
- Worldwide Distribution: USA and countries including: Canada, Turkey, India, Italy, France, the Netherlands, the UK, India, Spain, Greece, Saudi Arabia, Switzerland, Germany and China.
- Recall number
- Z-1008-2011
- Recall status
- Terminated
- CDRH record ID
- 88658
- FDA classified
- 2011-02-01
- Recall event number
- 54535