Open original FDA record ↗Recall record
- FDA identifier
- Z-1008-2015
- Company in FDA record
- Teleflex, Inc.
- FDA classification date
- Feb 6, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Connector mount cracks which may lead to a leak failure during use and necessitate immediate replacement in the breathing circuit. If leak exists and is left untreated, patient may be deprived of adequate anesthetic gases over a period of time and serious adverse health consequences may occur.
Source facts
- Action
- An Urgent - Field Safety Notice, dated December 4, 2014, was sent to end users that identified the product, problem, and action to be taken. Customers were requested to immediately check inventory, discontinue use and distribution, quarantine. Customers were to return the Acknowledgement form and request a return authorization number from their local Maquet representative for affected devices. A credit will be issued upon receipt of the returned affected product. For questions or for support concerning this issue, a local Maquet representative should be contacted at fieldactions@maquet.com.
- Initiated
- 2014-12-04
- Root cause
- Process control
- Terminated
- 2017-03-21
- Distribution
- Distributed in the states of FL, MI, WV, and KS.
- Recall number
- Z-1008-2015
- Recall status
- Terminated
- CDRH record ID
- 132281
- FDA classified
- 2015-02-06
- Recall event number
- 70127