Z-1008-2026: BVI Leos Laser and Endoscopy System. Model Number: OME6000U.

Recall record Z-1008-2026 for Beaver-Visitec International, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1008-2026
Company in FDA record
Beaver-Visitec International, Inc.
FDA classification date
Dec 31, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).

Source facts

Action
Beaver-Visitec International (BVI) notified consignees via emailed letter on 12/05/2025. Consignees were informed that the affected units can still be used. A BVI service technician will contact each consignee to schedule a visit. During this visit, the missing ferrites will be installed to restore the equipment to its fully compliant and validated configuration. BVI aims to complete the correction by December 31st, 2025.
Initiated
2025-12-05
Root cause
Process control
Distribution
US Nationwide distribution in the states of FL, MI, OK, TX, WI.
Recall number
Z-1008-2026
Recall status
Open, Classified
CDRH record ID
217578
FDA classified
2025-12-31
Recall event number
98076

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗