Open original FDA record ↗Recall record
- FDA identifier
- Z-1008-2026
- Company in FDA record
- Beaver-Visitec International, Inc.
- FDA classification date
- Dec 31, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Manufacturing deviation in which the ferrites specified for the USB cable assembly linking the Nurse Monitor touchscreen to the System Controller were not installed as intended. The omission of the ferrites on the USB communication cable results in non-conformance to the design specifications, which require two ferrites (one at each end of the cable).
Source facts
- Action
- Beaver-Visitec International (BVI) notified consignees via emailed letter on 12/05/2025. Consignees were informed that the affected units can still be used. A BVI service technician will contact each consignee to schedule a visit. During this visit, the missing ferrites will be installed to restore the equipment to its fully compliant and validated configuration. BVI aims to complete the correction by December 31st, 2025.
- Initiated
- 2025-12-05
- Root cause
- Process control
- Distribution
- US Nationwide distribution in the states of FL, MI, OK, TX, WI.
- Recall number
- Z-1008-2026
- Recall status
- Open, Classified
- CDRH record ID
- 217578
- FDA classified
- 2025-12-31
- Recall event number
- 98076