Skip to content

Public FDA record

Z-1009-03: Pulsar Max II Multiprogrammable Pacemaker, Model 1180

Recall record Z-1009-03 for Guidant Corp-Cpi Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1009-03
Company in FDA record
Guidant Corp-Cpi Division
FDA classification date
Jul 15, 2003
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Some of the pacemakers could have failed to pace due to an inadequately soldered electrical connection at an integrated circuit.

Source facts

Action
On June 23, 2003, sales representatives began retrieving any recalled pacemakers, which were not implanted. For implanted pacemakers, an 'Important Patient Management Information' letter, dated June 23, 2003, was sent to physicians recommending that pacemaker-dependent patients, with the potentially affected pacemakers, be scheduled for exams to check for proper pacemaker performance using a programmer.
Initiated
2003-06-23
Root cause
Other
Terminated
2004-07-02
Distribution
Nationwide in the United States and worldwide.
Recall number
Z-1009-03
Recall status
Terminated
CDRH record ID
27895
FDA classified
2003-07-15
Recall event number
26561

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗