Z-1009-2022: Accula SARS-CoV-2 Test, REF: COV4100

Recall record Z-1009-2022 for MESA BIOTECH, INC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1009-2022
Company in FDA record
MESA BIOTECH, INC
FDA record creation date
May 6, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

SARS-CoV-2 Test has an increased potential for false positive results, which may lead to a delay in correct diagnosis and treatment.

Source facts

Action
On 04/06/22, recall notices were mailed to customers. Customers were informed that the reliability of negative test results is not affected. Customers with affected devices were asked to do the following: 1. Discontinue use of any affected product immediately 2. All affected product must either be returned or discarded. Replacement product will be provided for all returned and discarded product upon receipt of Customer/Distributor Response Sheet. 3. Notify all impacted users in your facility. If you have shipped any of these lots outside of your facility, you must notify those customers or facilities of this voluntary recall. 4. Complete and return the Customer/Distributor Response Sheet, which can be emailed to USSD2.customercare@thermofisher.com Customers with questions are encouraged to contact the recalling firm at: +1-858-800-4929, or 1-800-955-6288, option 2.
Initiated
2022-04-06
Root cause
Material/Component Contamination
Distribution
US Nationwide distribution in the states of CA, TX, MA, NY, FL, NC, NJ, GA, CO, HI, PA, WA, NY, VA, MD, KY, WI, IL, PR, ND, MI, TN, AZ, UT.
Recall number
Z-1009-2022
Recall status
Open, Classified
CDRH record ID
193158
FDA record created
2022-05-06
Recall event number
90013

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗