Skip to content

Public FDA record

Z-1009-2024: Water Filter MAJ-2318. This MAJ-2318 water filter has been designed to be used with Olympus endoscope rep

Recall record Z-1009-2024 for Olympus Corporation of the Americas. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1009-2024
Company in FDA record
Olympus Corporation of the Americas
FDA classification date
Feb 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Water filter may have been assembled incorrectly, resulting in unfiltered tap water being used for reprocessing. This may expose a patient to a potentially contaminated device, presenting a risk of infection or sepsis.

Source facts

Action
On January 16, 2024, the firm notified customers of the issue via URGENT MEDICAL DEVICE FIELD CORRECTIVE ACTION letters. Customers were instructed to immediately assess all MAJ-2318 at their facility to check the model and lot number. Customers were given detailed instructions on how to identify affected product, including for MAJ-2318 installed in the OER machine. Please contact your Olympus customer service representative at 1-800-848-9024, option 2 to obtain a Return Material Authorization to return any affected product. Customers who have further distributed product should notify downstream accounts at once. If you require assistance with identifying affected product, please contact the Olympus Technical Assistance Center at 1- 800-848-9024, option 1. If you have questions about this recall, please contact the recalling firm at (647) 999-3203 or by e-mail at Cynthia.Ow@Olympus.com.
Initiated
2024-01-16
Root cause
Process control
Distribution
Domestic US distribution nationwide.
Recall number
Z-1009-2024
Recall status
Open, Classified
CDRH record ID
205621
FDA classified
2024-02-07
Recall event number
93792

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗