Issues identified: 1) Results vary by Instrument Platform; 2) Calibrator Lot-to-Lot Variability. Issues may lead to delayed or incorrect diagnosis, as well as delay of treatment or incorrect treatment.
Source facts
Action
Firm notified customers on November 21, 2025 via Urgent Medical Device Correction notices. Customers using the cobas e 402/801 and cobas e411 were instructed to immediately discontinue using the impacted lots outlined in the UMDC and switch to the new unaffected lot numbers listed in the communication.
Initiated
2025-11-21
Root cause
Process control
Distribution
US Nationwide distribution in the states of AZ, CA, FL, GA, IL, IN, ME, MI, MN, NC, NJ, NM, NY, OH, PA, PR, TN, TX, UT, WA, WV.
Recall number
Z-1009-2026
Recall status
Open, Classified
CDRH record ID
217618
FDA classified
2026-01-02
Recall event number
98103