Z-1010-2018: NaturaLyte Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is f

Recall record Z-1010-2018 for Fresenius Medical Care Renal Therapies Group, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1010-2018
Company in FDA record
Fresenius Medical Care Renal Therapies Group, LLC
FDA record creation date
Mar 14, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

The label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.

Source facts

Action
The firm, Fresenius Medical Care, sent an "URGENT - MEDICAL DEVICE RECALL" letter dated 10/4/2017 to its Consignees. The letter described the product, problem and actions to be taken. The Consignees are instructed to: " Immediately examine your stock to determine whether you have any Naturalyte Liquid Acid Concentrate from the lot listed. " If any product of this lot is found, discontinue use immediately. " Place all units in a secure, segregated area. " Contact your FMCRTG, LLC Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product. " Promptly fill out and return the attached Reply Form via fax to: 781-699-9796 ATTN: Quality Department or SCAN and email to: NOTIFYRA@fmc-na.com.' and "Route the letter to others in your organization that may have any of the affected product that is subject to this recall. Please note that the Naturalyte Liquid Acid Concentrate matching the Lot Number listed in the notification is the only lot subject to this recall. If you have any additional medical concerns or questions, please contact Medical Information and Communication: 855-616-2309 or Website: www.fresenius-medinfo.com.
Initiated
2017-09-29
Root cause
Release of Material/Component prior to receiving test results
Terminated
2018-08-09
Distribution
US Distribution to states of: CA, FL, KS, LA, and TX.
Recall number
Z-1010-2018
Recall status
Terminated
CDRH record ID
161727
FDA record created
2018-03-14
Recall event number
79236

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗