Skip to content

Public FDA record

Z-1010-2024: PulseSpray Infusion System devices including: PULSE SPRAY 5F/135/20, UPN: H787125004045, REF: 12500404; P

Recall record Z-1010-2024 for Angiodynamics, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1010-2024
Company in FDA record
Angiodynamics, Inc.
FDA classification date
Feb 7, 2024
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Rotating adaptor O-ring manufacturing defect impacting seal integrity which may lead to a leak during infusion of medication.

Source facts

Action
An Urgent Medical Voluntary Medical Device notices dated December 18, 2023 was issued to distributors and end users with instructions to immediately stop using affected product, remove said product from inventory and return to AngioDynamics, Inc. The notice should be forwarded to all site affected product may have been distributed to and recipients are asked to complete and return the Reply Verification Tracking Form. If affected product is located in your institution, please call AngioDynamics Customer Service at 1-800-772-6446 between 8:00 a.m. and 7:00 p.m. (Monday - Friday: Eastern Standard Time) to obtain a replacement or credit for your returned product.
Initiated
2023-12-18
Root cause
Process control
Distribution
United States (MD, LA & UT), Austria, Canada and Spain.
Recall number
Z-1010-2024
Recall status
Open, Classified
CDRH record ID
205352
FDA classified
2024-02-07
Recall event number
93744

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗