Open original FDA record ↗Recall record
- FDA identifier
- Z-1011-2012
- Company in FDA record
- Bio-Detek, Inc.
- FDA classification date
- Feb 13, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Lack of an adequate silicone seal around the periphery of the attenuator enclosure may allow fluid to enter.
Source facts
- Action
- Zoll Medical issued Urgent Device Correction notices dated 12/30/11 via UPS/Registered Mail. The letter described the issue and identified the affected product. ZOLL recommended that customers alert their staff of the issue, that the electrodes should be kept away from a fluid environment, to use caution in handling the electrodes if they are subjected to a fluid environment and maintain safe defibrillation practice, and seek a replacement if available. Also customers should contact Zoll 1-800-348-9011 for product replacement. The firm plans to replace the affected lots free of charge.
- Initiated
- 2011-12-30
- Root cause
- Process control
- Terminated
- 2012-12-11
- Distribution
- Worldwide Distribution -- US and Canada.
- Recall number
- Z-1011-2012
- Recall status
- Terminated
- CDRH record ID
- 106605
- FDA classified
- 2012-02-13
- Recall event number
- 60835