Z-1011-2015: BrightView model number: 882478 BrightView X model number: 882480 BrightView XCT model number: 882482 and

Recall record Z-1011-2015 for Philips Medical Systems (Cleveland) Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1011-2015
Company in FDA record
Philips Medical Systems (Cleveland) Inc
FDA classification date
Jan 26, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Unintended detector and gantry movement due to software issues.

Source facts

Action
Philips Healthcare sent an Urgent - Field Safety Notice dated November 4, 2014, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Philips is advising the customers/users to retain a copy of this Field Safety BY CUSTOMER I USER Notice with the equipment Instruction for Use, until further notice. Problem #1: Philips is advising the customer/user to perform the following workaround prior to the implementation of the field safety correction, to avoid contact between the detector and the head holder as describe above in Problem #1: ONLY if the system is positioned in a Relative-90 configuration with one of the detectors below the pallet level and the table in in the patient load position: Follow the sequence of operations prior to performing a study utilizing the head holder: - Deploy or store the x-ray panel if necessary; - Run the patient load PPM; - Install the head holder; and - Position the patient on the table and proceed with the study In addition, take the following preventative actions: " Monitor the patient during any and all system motions. " Know locations of the multiple E-Stop button(s) to halt all system motion. (Emergency Stop: where the system immediately stops all detector and gantry motions by disconnecting power to the motors that control system motion) Customers with questions were instructed to contact the Customer Care Solutions Certer at 1-800-722-9377. For questions regarding this recall call 440-483-2119.
Initiated
2014-11-12
Root cause
Software design
Terminated
2018-07-02
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Algeria, Argentina, Australia, Austria, Bahrain, Belgium, Brazil, China, Czech Republic, Denmark, Dominican Republic, Egypt, Finland, France, Gabon, Germany, Greece, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Lebanon, Lithuania, Malaysia, Mexico, Netherlands, New Zealand, Norway, Oman, Palestine, Panama, Philippines, Poland, Portugal, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Thailand, Turkey, Ukraine, United Arab Emirates, United Kingdom, United Kingdom, and Viet Nam.
Recall number
Z-1011-2015
Recall status
Terminated
CDRH record ID
131548
FDA classified
2015-01-26
Recall event number
69792

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗