Open original FDA record ↗Recall record
- FDA identifier
- Z-1011-2018
- Company in FDA record
- Zimmer Biomet, Inc.
- FDA record creation date
- Mar 14, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The affected products are labeled as a 62.5mm LEFT Femur; however, the contents are a Vanguard 70mm RIGHT Femur.
Source facts
- Action
- The firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL REMOVAL" letter dated October 3, 2017 to distributors and customers. The letter described the product, problem and actions to be taken. The customers were instructed to: 1. Review notification and ensure affected team member are aware of the contents 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of Attachment 1 Inventory Return Certification Form via digital copy to: corporatequallity.postmarket@zimmerbiomet.com within three (3) days. 4. Immediately return all affected product from your distributorship and affected hospitals within your territory along with a completed Attachment 1 Inventory Return Certification Form to Zimmer Biomet via SMS, via email to: rgarequest@zimmerbiomet.com-specify "RECALL" as RGA type. 5. Identify additional accounts Zimmer Biomet has not notified. 6. Retain a copy of your recall acknowledgement and product return forms for your records. Questions or concerns can be directed to the customer call center at 1-574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday.
- Initiated
- 2017-10-03
- Root cause
- Error in labeling
- Terminated
- 2018-08-23
- Distribution
- US Distribution to states of: FL, WI, MD, MO, and MI.
- Recall number
- Z-1011-2018
- Recall status
- Terminated
- CDRH record ID
- 161749
- FDA record created
- 2018-03-14
- Recall event number
- 79250