Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).
Source facts
Action
Aesculap sent an Important Correction and Removal and Field Safety Notification letter, dated December 17, 2014, to all affected customers. The letter identiifed the affected product, problem, and actions to be taken. Customers were requested to remove from inventory and contact Aesculap regarding return of the affected devices and acknowledgement form.
Initiated
2014-12-17
Root cause
Labeling False and Misleading
Terminated
2016-12-13
Distribution
Worldwide Distribution - US Nationwide
Recall number
Z-1012-2015
Recall status
Terminated
CDRH record ID
132226
FDA classified
2015-01-26
Recall event number
70086