Combination of CTVision with syngo RT Therapist / syngo RT Oncologist 4.3.SP1 automatic registration in Adaptive Targeting might result in wrong offset calculations. Applying this offset can lead to patient mistreatment. Cone Beam imaging is not affected by this problem.
Source facts
Action
A customer advisory notice, dated January 2, 2015, was sent to end users which identified the product, problem, and action to be taken. It was recommended that customers use only manual registration with the 6 degrees of freedom option disabled, which provides the correct offset values. A software update will be deployed once a solution is available.
Initiated
2015-01-02
Root cause
Software Design Change
Terminated
2015-09-10
Distribution
UT, WI, NY
Recall number
Z-1013-2015
Recall status
Terminated
CDRH record ID
132485
FDA classified
2015-01-26
Recall event number
70200