Z-1014-2022: ACL TOP 300 CTS/350 CTS/500 CTS/550 CTS/970 CL: Name Part Number 1.ACL TOP 300 CTS 00000280060 2. ACL T

Recall record Z-1014-2022 for Instrumentation Laboratory. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1014-2022
Company in FDA record
Instrumentation Laboratory
FDA classification date
Apr 28, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Updated and revised user instructions issued 7/14/22. Potential carryover issue may cause elevated quality control and sample results for heparin, apixaban or rivaroxaban when HemosIL Liquid Anti-Xa reagent is used in the same run with both HemosIL Liquid Antithrombin and HemosIL Q.F.A. Thrombin (Bovine), specific to instrument models that utilize a single reagent probe

Source facts

Action
*****UPDATE******PHASE 2 UPDATED - URGENT MEDICAL DEVICE CORRECTION This updated notification is intended to further clarify the Mandatory Customer Actions for the potential carryover issue identified on the following ACL TOP systems (previous URGENT MEDICAL DEVICE CORRECTION letter dated Feb 22, 2022), when running a specific reagent combination. E-mail communication to the Werfen Affiliate and International Group circulated on July 14, 2022. Subsequent investigations have shown that the carryover issue only has the possibility of occurring if QFA Fibrinogen, Liquid Antithrombin, and Liquid Anti-Xa reagent aspirations occur in a specific sequence. " Based on the above, we are revising the instruction to not run HemosIL Liquid Anti-Xa in the same batch as both HemosIL Liquid Antiithrombin and HemosIL Q.F.A. Thrombin (Bovine). Please take one of the following immediate workflows to mitigate this potential carryover issue. " Revised Instructions for Phase 2: 1. Use Separate instruments to run Liquid Anti-Xa and Liquid Antithrombin. 2. Batch Liquid Antithrombin.: Run Liquid Antithrombin tests in isolation (analyzer status starting and ending in a READY state). 3. Batch Anti-Xa, Apixaban and Rivaroxaban testing. Run the maintenance activity Enhanced Clean for Reagent Probe or Enhanced Clean for All Probes from the maintenance menu, then run Liquid Anti-Xa (heparin), apixaban, and rivaroxaban tests in isolation (analyzer status starting and ending in a READY state). Return the completed and signed Mandatory Response Tracking Form to: " Fax No.: 1-781-861-4207 Or E-mail: ra-usa.na@werfen.com For technical questions, please contact the Werfen Technical Support Center at 1-800-678-0710, Option # 1 (ACL Coagulation Products). Previous: PHASE 1: Werfen issued Urgent Field Safety Notification via email /certified mail on 2/22/22. Letter states reason for recall, health risk and action to take: Mandatory Customer Actions Based on the above, please take the following i
Initiated
2022-02-22
Root cause
Under Investigation by firm
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Belarus, Belgium, Bermuda, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Ecuador, El Salvador, France, Germany, Gibraltar, Great Britain, Greece, Guatemala, Guyana, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Japan, Korea, Kosovo, Kuwait, Lebanon, Lithuania, Luxembourg, Macau, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Norway, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Puerto Rico, Romania, Russia, S. Tome, Saudi Arabia, Serbia, Singapore, Slovakia, Spain, Switzerland, Taiwan, Tanzania, Thailand, Trinidad, Tobago, Tunisia, Turkey, United Arab Emirates, Uruguay, Venezuela, Vietnam.
Recall number
Z-1014-2022
Recall status
Open, Classified
CDRH record ID
192825
FDA classified
2022-04-28
Recall event number
89894

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗