The potential exists for external electrical noise to be processed as irregular pacing thereby masking potential asystole events.
Source facts
Action
The firm mailed an 'Urgent Medical Device Correction' letter to consignees on January 27, 2004 advising them of the external electrical noise problem and informing them to not use the Dinalink ILC1931 with patients with pace makers until a fix or replacement can be provided. By 5/17/2004 corrective replacement cables had been provided to all consignees.
Initiated
2004-01-27
Root cause
Other
Terminated
2009-10-05
Distribution
Product was distributed to 7 domestic hospitals in PA, SC, GA, FL, WI, SD AND WY.
Recall number
Z-1015-04
Recall status
Terminated
CDRH record ID
32943
FDA classified
2004-07-20
Recall event number
28997