Open original FDA record ↗Recall record
- FDA identifier
- Z-1015-2012
- Company in FDA record
- B. Braun Medical, Inc.
- FDA classification date
- Feb 29, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
If the IV set anti free flow clip is incorrectly inserted into the pump and the pump door forced closed, the clip catch on the inside of the pump door may break. If the clip catch is broken and the door opened, free flow protection is ensured. If IV set is then removed from pump without closing the roller clamp, the clip may not re-engage the tubing, leaving an open fluid path with free flow pote
Source facts
- Action
- B Braun sent a letter on August 16, 2011, to their customers, that detailed the proposed G03 implementation in response to reported customer complaints. The software upgrade with the door jam was agreed upon and the eight hospitals were upgraded in November 2011.
- Initiated
- 2011-07-12
- Root cause
- Device Design
- Terminated
- 2014-11-03
- Distribution
- Product was distributed to hospitals in CA and MO.
- Recall number
- Z-1015-2012
- Recall status
- Terminated
- CDRH record ID
- 107007
- FDA classified
- 2012-02-29
- Recall event number
- 60995