Open original FDA record ↗Recall record
- FDA identifier
- Z-1015-2013
- Company in FDA record
- CareFusion Corporation
- FDA classification date
- Mar 28, 2013
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
The recall was initiated because Carefusion has confirmed that when the PAS 2000 drawers due to inaccessability of drawers caused by system down time or lock-up.
Source facts
- Action
- Carefusion sent a Safety Alert Notification letter dated January 13, 2010 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to ensure back-up procedures were in place in case of system down-time or lock-up, including review of the back-up procedures with staff. The letter included an update to the PAS 2000 User Guide. The letterstates that a CareFusion field representative will be contacting customers to make arrangements to apply warning labels to the PAS 2000 electronics doors. Customers were instructed to complete and return the Response Card included. For questions contact your local field support representative ot the Advocacy department at customer.advocacy@carefusion.com.
- Initiated
- 2009-11-20
- Root cause
- Device Design
- Terminated
- 2013-03-28
- Distribution
- Worldwide Distribution - USA (Nationwide) including CANADA.
- Recall number
- Z-1015-2013
- Recall status
- Terminated
- CDRH record ID
- 86996
- FDA classified
- 2013-03-28
- Recall event number
- 53991