Some of the pacemakers could have failed to pace due to an inadequately soldered electrical connection at an integrated circuit.
Source facts
Action
On June 23, 2003, sales representatives began retrieving any recalled pacemakers, which were not implanted. For implanted pacemakers, an 'Important Patient Management Information' letter, dated June 23, 2003, was sent to physicians recommending that pacemaker-dependent patients, with the potentially affected pacemakers, be scheduled for exams to check for proper pacemaker performance using a programmer.
Initiated
2003-06-23
Root cause
Other
Terminated
2004-07-02
Distribution
Nationwide in the United States and worldwide.
Recall number
Z-1016-03
Recall status
Terminated
CDRH record ID
27966
FDA classified
2003-07-15
Recall event number
26561