Open original FDA record ↗Recall record
- FDA identifier
- Z-1016-2023
- Company in FDA record
- Future Diagnostics Solutions B.V.
- FDA classification date
- Jan 25, 2023
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Due to an issue of high coefficient of variability (CV) problem caused by the accuspheres inside the Immunoassay kits.
Source facts
- Action
- On 12/06/2022, the firm send a Recall Notice via email to customer informing them that The PTH kit 4K-IPT-00 with lot numbers: M78559, M78560 and M78636 can occasionally give a high duplicate %CV outside the recommended maximum of 15%CV. As a result of the failure in the duplicate CV requires a repeat run it may cause a delay in the patient result turnaround time or in system calibration depending on what samples are run and can lead to a wrong sample result. Actions to be taken by customers: -Immediately identify and quarantine the affected lot numbers. -Further instructions will be provided by the Future Diagnostics. For questions, send an email to fieldsafety@future-diagnostics.nl.
- Initiated
- 2022-12-06
- Root cause
- Under Investigation by firm
- Distribution
- Worldwide distribution. US states of AZ, CA, CO, LA, OH, PA, and TX; Austria, Denmark, France, Germany, Greece, Ireland, Italy, Netherlands, Spain, Sweden, and United Kingdom
- Recall number
- Z-1016-2023
- Recall status
- Open, Classified
- CDRH record ID
- 197520
- FDA classified
- 2023-01-25
- Recall event number
- 91344