Open original FDA record ↗Recall record
- FDA identifier
- Z-1017-2015
- Company in FDA record
- Covidien
- FDA classification date
- Feb 3, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
A manufacturing error resulted in the risk of incorrect proximal and distal balloon inflation port identification on the units. Units have been identified to have the distal balloon inflation port incorrectly labeled as proximal, and, the proximal balloon port incorrectly labeled as distal.
Source facts
- Action
- Consignees were sent an Urgent Product Recall letter dated 12/15/2014. The letter described the issue, identified the affected product, and the required actions. Affected product was to be returned directly to customers' Covidien Sales Rep; or customers were to contact Covidien Service at 1-800-716-6700 to arrange for product return. Questions can be directed to the Covidien Sales Rep or to Covidien Service at 1-800-716-6700. Customers were to complete and return the Verification Form along with the unused product as soon as possible.
- Initiated
- 2014-12-10
- Root cause
- Mixed-up of materials/components
- Terminated
- 2016-03-31
- Distribution
- Worldwide Distribution - US Nationwide and the countries Australia, Canada, Finland, France, Germany, Ireland, Italy, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
- Recall number
- Z-1017-2015
- Recall status
- Terminated
- CDRH record ID
- 132189
- FDA classified
- 2015-02-03
- Recall event number
- 70048