Skip to content

Public FDA record

Z-1017-2023: Invisalign Express 10 REF 8512 Invisalign System Comprehensive REF 9000 Invisalign System Moderate

Recall record Z-1017-2023 for Align Technology Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1017-2023
Company in FDA record
Align Technology Inc
FDA classification date
Jan 26, 2023
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

3D orthodontic planning software has a defect that leads to an issue where incorrect number of aligners may be produced (less or greater than the doctor requested) and incorrect packaging is provided.

Source facts

Action
On 12/05/2022, the firm initiated telephone calls with customer to inform them of a voluntary recall of selected Invisalign aligners. Starting on 12/09/2022, the firm sent an "URGENT Medical Device Correction" email to customers as a follow-up to a telephone call informing them that selected Invisalign aligners may have the potential clinical impact and therefore may result in unnecessary/unprescribed treatment due to software issue with the new Treatment Planning History feature within ClinCheck software. Customers are instructed: 1. Cease/stop using and dispose of the affected Invisalign case(s), including the aligners and packaging, and treatment PDF form-printed or online 2. Notify patients to discard the impacted aligners, if they have already been provided to them. The impacted product can be discarded as per the Doctor Instructions for Use or Patient Use and Care Instructions. As an immediate action, Align Technology has disabled the ClinCheck feature capability from the Treatment Plan History. For questions - contact CustomerSupport@aligntech.com or 888-822-5446.
Initiated
2022-12-05
Root cause
Software design
Distribution
U.S. Nationwide distribution in the states of AZ, CA, CO, FL, KY, MA, NJ, NY, OH, OK, and WI.
Recall number
Z-1017-2023
Recall status
Completed
CDRH record ID
197444
FDA classified
2023-01-26
Recall event number
91320

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗