Open original FDA record ↗Recall record
- FDA identifier
- Z-1018-2016
- Company in FDA record
- Sorin Group USA, Inc.
- FDA record creation date
- Feb 25, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The CDI System 500 monitor displays the "H/S DISCONNECT AT CUVETTE" error message when the CDI HIS Cuvette does not make a proper connection with the CDI H/S Probe. When this occurs, blood parameter values for HCT, Hgb and S02 do not display.
Source facts
- Action
- A Safety Alert letter dated January 13, 2015 via registered mail to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to check the communication of the Cuvette prior to initiating Cardiopulmonary Bypass (CPS). The Terumo Safety Alert did not request removal or return of the product. The instructions that Sarin has communicated to the customers is identical to the information from Terumo. Acknowledgement of the letter and understanding of the contents is required for all customer responses.
- Initiated
- 2016-01-19
- Root cause
- Component design/selection
- Terminated
- 2016-10-24
- Distribution
- US Nationwide Distribution and Canada.
- Recall number
- Z-1018-2016
- Recall status
- Terminated
- CDRH record ID
- 143117
- FDA record created
- 2016-02-25
- Recall event number
- 73126