Skip to content

Public FDA record

Z-1018-2022: BD Synapsys Microbiology Informatics Solution, Catalog Number 444150, Software Version 4.10

Recall record Z-1018-2022 for Becton Dickinson & Co.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1018-2022
Company in FDA record
Becton Dickinson & Co.
FDA record creation date
Apr 29, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

When updating BD Kiestra customers to BD Synapsys version 4.10, the custom antibiotic codes set by the customer are overwritten with the BD default antibiotic codes.

Source facts

Action
A MEDICAL DEVICE CORRECTION notification letter dated 3/9/22 was distributed to customers. PLEASE TAKE THE FOLLOWING ACTIONS: 1. Ensure the contents of this notification are read and understood by those within your organization. 2. BD will be contacting you within one week to request remote access to your database to confirm your configuration and evaluate potential impact. If BD determines that your facility has an affected configuration, BD will work with your facility to resolve the issue. In the interim, you many continue use of the product until BD contacts your facility. 3. Complete the attached Customer Response Form and return to the BD contact noted on the form so that BD may acknowledge your receipt of this notification per FDA requirements. 4. Report any adverse health consequences experienced with the use of this product to BD. Events may also be reported to the FDA's MedWatch Adverse Event Reporting program via: Web: MedWatch website at www.fda.gov/medwatch Phone: 1-800-FDA-1088 (1-800-332-1088) Mail: MedWatch, HF-2, FDA, 5600 Fisher s Lane, Rockville, MD 20852-9787 Actions Taken by BD: 1. BD will review configuration and evaluate potential impact for BD Kiestra customers. 2. Corrective actions have been initiated to prevent recurrence of the identified root cause. If you require further assistance, please contact:
Initiated
2022-03-09
Root cause
Under Investigation by firm
Terminated
2024-01-19
Distribution
Worldwide distribution US Nationwide distribution in the states of CA, IL, MI, NY, OH, SC, and TN. The countries of Austria, France, Germany, Netherlands, Sweden.
Recall number
Z-1018-2022
Recall status
Terminated
CDRH record ID
192826
FDA record created
2022-04-29
Recall event number
89896

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.