Z-1018-2026: VANTAGE TITAN 3T Model MRT-3010/MEXL-3010

Recall record Z-1018-2026 for Canon Medical System, USA, INC.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1018-2026
Company in FDA record
Canon Medical System, USA, INC.
FDA classification date
Jan 13, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Source facts

Action
On 12/19/2025, the firm sent via Adobe DocuSign an "Urgent: Medical Device Correction" letter to customers informing them that It has been discovered that in the superconducting magnet used in the applicable system, if there is slow leakage, over a long period of time, of refrigerant gas from the superconducting magnet, air ingress into the helium vessel via the leak point may result in ice forming. If this is left unchecked, the worst-case scenario is that the helium gas will not be able to escape through the designed exhaust path, and the helium tank could eventually rupture, releasing helium gas into the MR scan room. For safety management of the superconducting magnet, the status of the magnet is checked during maintenance inspections and continuously monitored by a remote monitoring system. However, in response to the above report from the manufacturer of the superconducting magnet, an inspection to check for slow gas leaks in the superconducting magnet used in systems installed at our customer sites will be provided as a repair service. Note that there have been no reports of health hazards related to this incident. Customer are instructed to: The system can continue to be used. However, please ensure there are adequate evacuation routes from the scan/imaging room and the control room. Additionally, do not press the quench button in cases other than emergencies. A representative of Canon Medical Systems USA will contact you to schedule an inspection. If you have any questions regarding this matter, please contact your service representative. Share the contents of letter with all users and reviewing radiologist as well as clinical engineering or biomedical group at their facility. In addition, posted the Recall Letter on or near potentially affected systems. For any questions regarding this letter please contact Orlando Tadeo, Director Regulatory Affairs, at 800-421-1968 or otadeo@us.medical.canon or your Canon service representative at 800-521-1968.
Initiated
2025-12-19
Root cause
Under Investigation by firm
Distribution
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Recall number
Z-1018-2026
Recall status
Open, Classified
CDRH record ID
217629
FDA classified
2026-01-13
Recall event number
98112

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗