Skip to content

Public FDA record

Z-1019-05: HomeChoice PRO Automated Peritoneal Dialysis Systems; Manufactured for Baxter Healthcare Corporation, Dee

Recall record Z-1019-05 for Baxter Healthcare Renal Div. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1019-05
Company in FDA record
Baxter Healthcare Renal Div
FDA classification date
Jul 19, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The screws that secure the Power Entry Module (PEM) to the base of the HomeChoice device are not reliably grounded to the device's ground system, posing an elcetric shock hazard.

Source facts

Action
Baxter sent Urgent Device Correction letters worldwide to their customers on 6/21/05 and home patients on 6/22/05. The accounts were informed of teh risk of electrical shock from screws used to secure the Power Entry Module to the back of the HomeChoice system, and were instructed to always connect the power cord to the power entry module first, before plugging the power cord into a grounded electrical outlet. The users were advised to use caution to avoid contact the power entry module screws while powering on or powering off the device, and were informed that a new mounting bracket that providews more reliable grounding was installed in instruments processed through the Baxter service organization as of May 2005. All remaining instruments will receive the new mounting bracket by August 2006, and Baxter will contact the accounts to arrange for the upgrade. Any questions were directed to Baxter Global Technical Services at 1-800-553-6898, prompt 1.
Initiated
2005-06-21
Root cause
Other
Terminated
2007-12-11
Distribution
Nationwide, including Puerto Rico, and internationally to Argentina, Austria, Australia, Belgium, Canada, Chile, China, Colombia, Denmark, Ecuador, France, Germany, Guatemala, Hong Kong, Ireland, India, Italy, Korea, Mexico, Malaysia, Nicaragua, the Netherlands, New Zealand, Peru, the Philippines, Portugal, Singapore, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkey, the United Kingdom, Uruguay and Venezuela.
Recall number
Z-1019-05
Recall status
Terminated
CDRH record ID
39446
FDA classified
2005-07-19
Recall event number
32251

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.